Automatic steam sterilizer GPa-10 PZ
Паровые стерилизаторы (Автоклавы)

Automatic steam sterilizer GPa-10 PZ

Automatic steam sterilizer GPa-10 PZ Designed for sterilization with saturated water steam under excess pressure of medical products made of metal (surgical instruments, etc.), rubber (surgical gloves, etc.), glass (dishes, etc.), plastic, as well as dressings and ligature suture materials, textiles (surgical linen, etc.) in packaged and unpackaged form. The sterilizer has a filter for bactericidal air purification, entering the sterilization chamber during the pressure equalization stage. To ensure effective removal of air from the sterilization chamber and to ensure vacuum drying, the sterilizer is connected to the water supply and sewerage system. Scope of application: Dental and gynecological offices, ophthalmological operating rooms, medical institutions, laboratories. Benefits: possibility of choosing sterilization modes; the ability to sterilize medical instruments of complex shape, hollow inside, with small holes; pre-vacuum, vacuum drying (carried out using a condenser) and rapid sterilization cycle; effective sterilization when starting from a cold state; full automation of the sterilization process; microprocessor control provides constant and most accurate control of the parameters and functions of the sterilizer; the ability to change the parameters of the sterilization mode (temperature, time), parameters of the pre-vacuum stage, drying time allows users to flexibly control all parameters of the sterilization process, depending on the products being sterilized; minimum sterilization cycle - 45 minutes; all elements of the sterilizer vessel, except for seals, pipelines and fittings, are made of corrosion-resistant steel; the sterilizer is equipped with an interface connector for connecting to a computer; non-volatile memory. Manba: https://www.nv-lab.ru/catalog_info.php?ID=1982

Price on request
  • Parameter: Value
  • Camera type: rectangular
  • Useful chamber volume, l: 10
  • Maximum operating pressure, MPa: 0.22
  • Pressure maintenance accuracy, MPa: ±0.01
  • Volume of distilled water poured in when starting the sterilizer for the first time, l: 2.75
  • Residual moisture content of sterile textile materials, %, not more than: 1
  • Internal dimensions of the chamber (W × D × H), mm: 200 × 280 × 180
  • External dimensions (W × D × H), mm: 453 × 718 × 396
  • Weight, kg: 44±10%
  • Power consumption, kW, no more: 1.8
  • Power, V/Hz: 220/50
  • Specified temperature conditions, °C: 134 °C – 5 min;
  • 121 °C – 20 min;
  • 126 °C – 10 min
  • Sterilizer heating time, min, no more than: 30
  • Maximum temperature deviation in the loaded chamber from the nominal value, °C, no more than: ±1
  • Number of shelves, standard/max., pcs.: 2/3
  • Shelf dimensions, mm: 178 × 260
  • Corrected sound power level, dBA, no more than: 70
  • Continuous operation time per day, h, not less: 16
  • MTBF, cycles, not less: 1000
  • Average service life, years: 10