Portable conductivity meter ECOSTAB EC242
Portable conductivity meter ECOSTAB EC242 Portable conductivity meter with GLP (NLP), memory and USB output. Designed for laboratory measurement of specific electrical conductivity (SEC) and work in field conditions. The model is equipped with various intelligent functions and has a durable design. The case has a dust and moisture protection rating of IP57, is equipped with a silicone protective cover for the sensor connectors and a cover plate that protects the tool from shock and fits well in the hand. Large LCD display displays measured parameters and temperature, indication of stable readings and completed calibrations. The screen is protected by an acrylic coating that is resistant to scratches and water ingress. The folding stand allows the device to be used as a tabletop device. Standard BNC connector compatible with LabSen® professional electrodes for a variety of applications. Built-in self-diagnostic function and graphical calibration instructions on the device screen. Automatic Temperature Compensation (ATC) ensures accuracy at different temperatures. Supplied with a case and a starter kit for analysis (electrode, electrode mount, flash drive with software, USB cable, conductivity standards (84 µS, 1413 µS, 12.88 mS) 50 ml each, 3 AAA batteries, screwdriver for replacing batteries). Applications: Laboratory analysis. Drinks and juices. Water supply. Ecological monitoring. Treatment facilities. Clean/drinking/waste water. Manba: https://www.nv-lab.ru/catalog_info.php?ID=7927
- Parameter: Value
- Measured parameter: Electrical conductivity (SEC) / Temperature
- Measuring ranges:
- - UEP, µS/cm;
- - temperature, °C: from 0.8 to 200000; *
- from 0 to 50
- Resolution:
- - UEP, µS/cm;
- - temperature, °C: 0.01;
- 0.1
- Absolute measurement error:
- - UEP, %;
- - temperature, °C: ±1; **
- ±1
- Cell constant, cm-1: 0.1 / 1 / 10
- Calibration: Automatic from 1 to 3 points
- Functions: Calibration reminder and calibration time check: Yes
- Temperature compensation, °C: Automatic (ATC) from 0 to 50 (reference temperature 15 – 30)
- Display: LCD with backlight
- Auto Hold: Yes
- Stable measurement indication: smiley icon
- Self-diagnosis: Yes
- GLP data management (NLP), data storage (number, parameter and temperature measurement value, measurement data and time): 500 records
- Function: date and time, date and time storage: Yes
- USB interface: Yes
- Dust and moisture protection degree: IP57
- Power: AAA batteries × 3 pcs.
- Dimensions, mm
- - tester;
- - with case: 88 × 170 × 33;
- 360×270×76
- Weight
- - tester, g;
- - with case, kg: 313;
- 1.6