Mon-Fri 09:00-18:00
EXON IMPEX Modern technologies for education and business
Conductivity meter HANNA HI 98311 Dist5 (pocket)
Кондуктометры

Conductivity meter HANNA HI 98311 Dist5 (pocket)

Conductivity meter HANNA HI 98311 Dist5 (pocket) DIST-5 is a pocket-sized and accurate device for monitoring conductivity and salinity. Features and Benefits: Waterproof housing. Open temperature sensor for fast response. Low range of electrical conductivity from 0 to 3999 µS/cm and salt content from 0 to 2000 parts per million (mg/l). HOLD button to freeze the current readings on the display. The BEPS function alerts the user when the battery level is low, which may negatively affect the tester readings. The conductivity meter-salin meter DIST 5 (HI 98311) can operate in conductivity mode and in salt content (TDS). The conversion factor from conductivity units to salt content units (TDS factor) can be changed in the range from 0.45 to 1.00 (standard – 0.50). The display simultaneously shows the temperature value of the sample being measured. The temperature coefficient can be set from 0.0 to 2.4%/°. This tester has the following features: replaceable graphite electrode, variable salinity conversion factor, temperature measurements in °C or °F, automatic temperature compensation (ATC) with adjustable β, battery level indicator, stability indicator, automatic shutdown and automatic one-point calibration. The conductivity cell is easy to replace if necessary. Due to the low degree of contamination by salt deposits from the samples, the graphite electrode provides greater accuracy in conductivity measurements. All these features are enclosed in a waterproof case. This 3-in-1 tester is unrivaled in measuring electrical conductivity, salinity and temperature. Manba: https://www.nv-lab.ru/catalog_info.php?ID=1397

Price on request
  • Parameter: Value
  • Conductivity range, µS/cm: from 0 to 3999
  • Salinity range: 0 to 2000 ppm (mg/l)
  • Temperature range, °C: 0.0 to 60.0 (32.0 to 140.0 °F)
  • Conductivity resolution, µS/cm: 1
  • Salinity resolution: 1 ppm (mg/l)
  • Temperature resolution, °C: 0.1 (0.1 °F)
  • Conductivity/salinity accuracy: ±2% FS
  • Temperature accuracy, °C: ±0.5 (±1 °F)
  • Conductivity/Salinity Calibration: Automatic, Single Point: 1413 µS/cm, 1382 ppm (mg/L)
  • Conversion factor to salt content: from 0.45 to 1.00
  • Temperature compensation: automatic, with β adjustable from 0.0 to 2.4%/°C
  • Electrode/probe: HI 73311 replacement graphite conductivity/salinity electrode (included)
  • Battery type/durability: 1.5V (4 pcs)/approximately 100 hours of continuous use, auto shuts off after 8 minutes of non-use
  • Environment: 0 to 50°C (32 to 122°F); relative humidity max. 100% no condensation
  • Dimensions, mm: 163 × 40 × 26 mm (6.4 × 1.6 × 1.0'')
  • Weight, g: 100 (3.5 oz)
  • Ordering information: HI 98311 (DiST®5) tester comes with HI 73311 conductivity/salinity electrode, HI 73128 electrode removal tool, batteries and instructions

Solution selection

We will recommend a setup for a classroom, hall or laboratory

The request is saved to the database so a manager can process it later.