Ion selective electrode ELIT-071NO2 (glass)
Ионоселективные электроды

Ion selective electrode ELIT-071NO2 (glass)

Ion selective electrode ELIT-071NO2. Designed to determine the concentration of nitrite ions in meat products, as well as brines and curing mixtures using the potentiometric method. Manba: https://www.nv-lab.ru/catalog_info.php?ID=1529

Price on request
  • Feature: Parameter
  • Determination range of activity (concentration) of nitrite ions, рСNO₂: 5.00-2.52
  • Linear range for determining the activity (concentration) of nitrate ions, рСNO₂: 5.00-2.52
  • Electrical resistance of the electrode at a temperature of (20±5)°C, MOhm: does not exceed 100
  • Slope of the linear part of the electrode calibration characteristic at solution temperature (20±5)°С, mV/рСNO₂: 53±6
  • Deviation of the calibration characteristic from linearity, mV: no more than ±6
  • Temperature range of the analyzed solution, °C: 5-45
  • Permissible pH range of the analyzed solution, units. pH: 4.5-8.0