Low-frequency physiotherapy device "Electroson-Br"
The low-frequency physiotherapeutic device "Electroson-Br" is intended for carrying out electrosleep procedures - a method of electrotherapy that involves influencing the central nervous system with pulsed currents of low frequency, low strength, and a rectangular configuration. The device is intended for use in physiotherapy rooms of clinics, hospitals, medical institutions. ATTENTION! The device is contraindicated for persons with an implanted device (for example, a pacemaker). Under the influence of electrosleep therapy, the following pronounced therapeutic effects appear: sedative and tranquilizing (emotional excitability decreases, mood improves, natural sleep normalizes); analgesic (due to stimulation of the opioid system of the brain); hypotensive; hemodynamic (the central and autonomic regulation of the cardiovascular system is rebuilt without negative changes in the coronary and cerebral circulatory system); hormonal (pronounced stimulation of the pituitary gland function); immunomodulatory; metabolic-trophic (influences the central mechanisms of various metabolic processes in the body (carbohydrate, lipid, purine, oxygen transport function of the blood improves); regenerative (in case of various damaging processes, it stimulates regeneration processes). The Electroson-Br device is made on a modern element base, has an ergonomic control panel with the ability to adjust the amplitude of the pulse current and the additional constant component of the patient's current, and set the pulse repetition rate. The device allows you to set the duration of the procedure up to 99 minutes and display the countdown. After the procedure time has elapsed, the patient's current is automatically switched off. The device kit includes: extension cable, cable with electrode mask for electrosleep 62 mm, cable with electrode mask for electrosleep 68 mm The device is safe and easy to use. Manba: https://www.nv-lab.ru/catalog_info.php?ID=2173
- – amplitude of the pulse current on the load from 0
- up to 5 kOhm (0 – 10) mA
- - additional constant
- load component from 0 to 5 kOhm (0 – 1) mA
- – pulse repetition rate 5,10,20,40,80,100 and 160 Hz
- – pulse duration 0.5 ms ± 20%
- – duration of the rise and fall of pulses is no more than 50 μs
- -procedural timer allows you to set the duration
- procedures up to 99 min. with discreteness of 1 min.
- -it is possible to change the value of current, frequency, time
- procedures after starting the procedure
- - the device has protection that turns off the output current,
- if its total value for reasons of malfunction
- device will exceed the set current value by more than
- than 20%
- – power consumption no more than 10 VA
- – overall dimensions (mm) 260x210x100
- – electrical safety (protection class) II
- – the weight of the device without a set of accessories is no more than 1.5 kg
- Apparatus
- intended for use in the following conditions:
- – ambient temperature from 10 to 35 °C;
- – relative air humidity up to 80% at temperature
- 25 °C;
- – atmospheric pressure from 86 to 106 kPa (from 650 to 800 mm Hg);
- – supply voltage (220±22)
- B with a frequency of 50±0.5 Hz.